Ludlum Model 14C Survey Meter

The Ludlum Model 14C survey meter was designed to meet nuclear medicine requirements. The Model 14C is similar to the Model 3, but with an internal energy compensated G-M detector. It can be used with one or more external GM or scintillation detectors for alpha, beta, or gamma detection up to 0-200 mR/hr.  This additional internal detector creates fifth high range of 0-2000 mR/hr in addition to the four ranges covered by the external probe.  It is used in hosptials, universities, power plants and regulatory facilities. Standard scale is 0-2 mR/hr with five counting ranges: x0.1, x1, x10, x100, and x1000. (Dual scale cpm & mR/hr are available on request). Two “D” cells provide over 200 hours of operation. The Model 14C is supplied as shown with built-in speaker. 

Try the new lighter weight Model GSM-506, featuring auto-ranging, integration, and all-microprocessor circuitry.

COMPATIBLE DETECTORS: G-M, scintillation
METER DIAL:Typically 0 - 2 mR/hr and cpm, BAT TEST (others available )
MULTIPLIERS: X0.1, X1, X10, X100, X1000
LINEARITY: Reading within plus or minus 10% of true value with detector connected
CONNECTOR: Series "C" (others available )
INTERNAL DETECTOR: Energy compensated G-M (used with X1000 scale only)
ENERGY RESPONSE: Within plus or minus 15% of true value between 60 keV - 3 MeV (internal detector only )
THRESHOLD: 30 mV plus or minus 10 mV
AUDIO: Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet )
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided )
HIGH VOLTAGE:900 volts
THRESHOLD: 30 mV plus or minus 10 mV
RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
RESET: Push-button to zero meter
POWER: 2 each "D" cell batteries (housed in sealed compartment that is externally accessible )
BATTERY LIFE: Typically greater than 1000 hrs.(battery condition can be checked on meter )
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 6.5" (16.5 cm)H X 3.5" (8.9 cm)W X 8.5" (21.6 cm)L
WEIGHT: 3.5 lbs (1.6 kg) including batteries

Note that the Ludlum Model 14C Survey Meter houses an internal energy-compensated G-M detector with limitations listed in the description. External suggested detectors are listed below and are not included with the survey meter price. An additional charge will be added for each detector ordered.  

Recommended external probes for the Ludlum Model 14C:

For
Alpha
Detection

For
Beta
Detection

For
Gamma
Detection

For Alpha/beta/gamma Detection

Model 44-7
Model 44-9
Model 44-88
Model 44-89
Model 44-94

Model 44-6
Model 44-7
Model 44-9
Model 44-38
Model 44-88
Model 44-89
Model 44-94

Model 44-2
Model 44-3
Model 44-6
Model 44-7
Model 44-9
Model 44-38

Model 44-7
Model 44-9
Model 44-88
Model 44-89
Model 44-94

Ludlum Model 14C Common Meter dials:

 202-859
0 - 5k cpm; 0-2 mR/hr
For Any Detector

 202-608
0-6.6k cpm; 0-2 mR/hr
For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94

 202-664
0-6.6k cpm; 0-20
mSv/h
For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94

 


 

202-330
0-4k cpm; 0-2 mR/hr
For Model 44-7
 

 202-676
0-4k cpm; 0-20
mSv/h
For Model 44-7

202-084
0-2 mR/hr
For Model 44-6; 44-38

 


202-379
0-20
mSv/h
For Model 44-6; 44-38

202-241
0-2 mR/hr; 0-2.4k cpm
For Model 44-6; 44-38

202-546
0-20
mSv/h; 0-2.4k cpm
For Model 44-6; 44-38

Accessories

Replacement Parts

Check Source
Source Holder
Overload Indicator
Scaler Option
Shoulder Strap
Headset
Standard Carrying Case
Air and Water Tight Carrying Case

Cables
Meter Bezel
Meter Movement
Handle
Probe Clip

Ludlum Model 14C Manual        Ludlum Model 14c Response Curve

 DS 051-014 Model 14C Survey Meter  
 DS 051-014-7 Model 14C Survey Meter with Model 44-7 probe  
 DS 051-014-9 Model 14C Survey Meter w/ Model 44-9 probe  
 DS 051-014-3/9 Model 14C Survey Meter w/ 44-9 and 44-3 NaI probe  

 

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific    tel (310) 589 0601   fax (310) 919-2981,   Email

contact us: en espaņol, auf Deutsch
Open 8:00 AM - 5:00 PM Pacific Standard Time