• Built-In 1" X 1" NaI(Tl) Gamma Scintillator
  • 4 Ranges
  • Built In Audio
  • Total Counting Range from 0 - 3000 ĩR/hr

PART NUMBER:48-1610

INDICATED USE: Low level (microR) gamma survey
DETECTOR: 1" X 1" sodium iodide (NaI)Tl scintillator
SENSITIVITY: Typically 175 cpm/microR/hr (137Cs gamma )
ENERGY RESPONSE: Energy dependent
METER DIAL: 0 - 3 microR/hr, BAT TEST (others available )
MULTIPLIERS: X1, X10, X100, X1000
LINEARITY: Reading within plus or minus 10% of true value
AUDIO: Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet )
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided )
RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
RESET: Push-button to zero meter
POWER: 2 each "D" cell batteries (housed in sealed compartment that is externally accessible )
BATTERY LIFE: Typically 600 hours with alkaline batteries (battery condition can be checked on meter )
METER: 2.5" (6.4cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 7.8"(19.8 cm)H X 3.5"(8.9 cm)W X 8.5"(21.6 cm)L
WEIGHT: 4.5 lbs(2.1 kg) including batteries

Waterproof Pelican Cases Available

Common Meter Dial

 

202-004
0 - 3
mR/hr

202-419
0 - 0.03
mSv/h

 

 

Accessories

Replacement Parts

Check Source
Headset
Standard Carrying Case
Air and Water Tight Carrying Case

Meter Bezel
Meter Movement
Handle
Detector Assembly

 

Manual

 

Instrumentation

Ludlum Dosimetry Shielding Portable Fume Hoods Specials

Hepa Vacuums

Air Sampling

Beakers

 Sources

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy SEARCH

Home

to order:    tel (888) 368 9118 or (310) 589 0601   fax (888) 440-1319,   Email

Direct Scientific / 124 San Tropez Ct. / Laguna Beach, California 92651  
contact us: en espaņol, auf Deutsch