Model 2241-3 Digital Survey Meter
Digital LED Display with Backlight
PART# :48-2864
INDICATED USE: General purpose survey,
gross counting
COMPATIBLE DETECTORS: G-M, proportional, scintillation
CONTROL SWITCH: A 5 position rotating switch to allow for quick
change between four detector setups
CONNECTOR: Series "C" (others available on request)
AUDIO: Built in unimorph speaker with ON/OFF switch (greater than
60 dB at 2 feet)
ALERT/ALARM: Indicated by enunciator on display and audible tone
DISPLAY: 4 digit LCD display with 0.5"(1.3cm) high digits, separate
enunciators for display units, alert, alarm, low battery, detector overload,
counting overflow, and scaler counting
BACKLIGHT: Push-button to activate
RATEMETER: Can display in R/hr, Sv/hr, cpm, or cps when control
switch is in RATEMETER position
DISPLAY RANGE: Auto ranging from 0.0 microR/hr - 9999 R/hr; 0.000
microSv/hr - 9999 Sv/hr; 0 cpm - 999k cpm; or 0 cps - 100k cps
LINEARITY: Reading within plus or minus 10% of true value with
detector connected
SCALER: Activated by push-button in handle (count time adjustable
from 1 - 9999 seconds in 1 second intervals)
CALIBRATION CONTROLS: Accessible from front of instrument (protective
cover provided)
HIGH VOLTAGE: Four independent controls; each adjustable from 200 -
2500 volts
DISCRIMINATOR: Adjustable from 2 - 100 mV
OVERLOAD: Indicated by OVERLOAD on display (adjustable depending
on detector selected)
RESET: Push-button to zero display, acknowledge and/or reset alarm
POWER: 2 each "D" cell batteries (housed in sealed compartment
that is externally accessible)
BATTERY LIFE: Typically 200 hours with alkaline batteries (24
hours when low battery is first indicated on display)
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel
paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 6.5"(16.5cm)H X 3.5"(8.9cm)W X 8.5"(21.6cm)L
WEIGHT: 3.5 lbs (1.6kg) including batteries
The following parameters are adjusted by
removable switchboard. Once initial setup and calibration have been
completed the switchboard can be removed without affecting normal operation
of the instrument (Model 2241-2 may be ordered without switchboard)
BACKLIGHT "ON" TIME: 5, 30, 60, 90, 180, or 240 seconds for the
backlight to stay "ON" when activated.
SET MINIMUM DISPLAY: Allows lower limit of auto-ranging display to be
fixed. For example, display can be set to not show values less than 1 R/hr.
RS-232 DATA DUMP MODE: Enables or disables dump mode to RS-232 port.
When enabled data will be dumped every 2 seconds
RS-232 DETECTOR SETUP MODE: Allows for detector parameters to be
input by RS- 232 port
BAUD RATE: 150, 300, 600, 1200, 2400, 4800, 9600, or 19200 bps.
DEAD TIME: Adjustable from 0 - 9999 seconds
CALIBRATION CONSTANT: Adjustable from 0.001 - 280 X 109
counts/display unit
DISPLAY UNITS: Can read in R/hr, Sv/hr, cpm, or cps
TIME BASE: Can read in seconds or minutes
AUDIO DIVIDE: 1, 10, 100, or 1000 events-per-click
RESPONSE:
Fixed: FAST response time adjustable from
approximately 2 - 50 seconds. SLOW response is 5 times
fast setting
Variable: Dependent on number of counts present.
Typical times FAST - 4 - 25 seconds, SLOW - 4-60
seconds
(All times stated from 10% to 90% of final reading)
RATEMETER ALERT/ALARM: Set at any point corresponding to pre-selected
ratemeter range
SCALER ALARM: Adjustable from 1 - 999999 counts
SCALER COUNT TIME: Adjustable from 1 - 9999 seconds
![]()
|
For |
For |
For |
For Alpha/beta/gamma Measurements |
|
Model 43-1 |
Model 43-20 |
Model 44-2 |
![]()
Model 2241 Series Digital Display

![]()
|
Accessories |
Replacement Parts |
|
Check Source |
Cable |
![]()
![]()
|