Model 2241-3i Digital Scaler / Ratemeter
-
Up
to 3 External Detectors and Internal Detector
- Digital LED Display with Backlight
- Auto-Ranging
- Built-In Scaler
- Alert and Alarm
- Utilizes G-M, Proportional, and
Scintillation Detectors
- RS-232 Data Dump
PART #:48-3358
INDICATED USE: General purpose survey, gross counting
COMPATIBLE DETECTORS: G-M, proportional, scintillation
CONTROL SWITCH: A 5 position rotating switch to allow for quick change
among 3 external detectors or the internal G-M.
INTERNAL DETECTOR: Energy compensated G-M
INTERNAL DETECTOR LINEAR RANGE: 1mR/hr to 10 R/hr
ALERT/ALARM: Indicated by annunciator on display and audible tone
DISPLAY: 4 digit LCD display with 0.5"(1.3cm) high digits, separate
annunciators for display units, alert, alarm, low battery, detector overload,
counting overflow, and scaler counting
BACKLIGHT: Push-button to activate
RATEMETER: Can display in R/hr, Sv/hr, cpm, or cps when control switch is
in RATEMETER position
DISPLAY RANGE: Auto ranging from 0.0 µR/hr - 9999 R/hr; 0.000 µSv/hr -
9999 Sv/hr; 0.00 cpm - 999k cpm; or 0.00 cps - 100k cps
SCALER: Activated by push-button in handle (count time adjustable from
1 - 9999 seconds in 1 second intervals)
HIGH VOLTAGE: Four independent controls; each adjustable from 200 - 2500
volts
DISCRIMINATOR: Adjustable from 2 - 100 mV
OVERLOAD: Indicated by OVERLOAD on display (adjustable depending on
detector selected)
BATTERY LIFE: Typically 200 hours with alkaline batteries (24 hours
when low battery is first indicated on display)
SIZE: 6.5"(16.5cm)H X 3.5"(8.9cm)W X 8.5"(21.6cm)L
WEIGHT: 3.5 lbs (1.6kg) including batteries
The following parameters are adjusted for each
detector by removable switchboard. Once initial setup and calibration have been
completed the switchboard can be removed without affecting normal operation of
the instrument (Model 2241-3i may be ordered without switchboard)
BACKLIGHT "ON" TIME: 5, 30, 60, 90, 180, or 240 seconds for the backlight
to stay "ON" when activated.
SET MINIMUM DISPLAY: Allows lower limit of auto-ranging display to be
fixed. For example, display can be set to not show values less than 1
R/hr.
RS-232 DATA DUMP MODE: Enables or disables dump mode to RS-232 port. When
enabled data will be dumped every 2 seconds
RS-232 DETECTOR SETUP MODE: Allows for detector parameters to be input by
RS- 232 port
BAUD RATE: 150, 300, 600, 1200, 2400, 4800, 9600, or 19200 bps.
DEAD TIME: Adjustable from 0 - 9999 seconds
CALIBRATION CONSTANT: Adjustable from 0.001 - 280 X 109
counts/display unit
DISPLAY UNITS: Can read in R/hr, Sv/hr, cpm, or cps
TIME BASE: Can read in seconds or minutes
AUDIO DIVIDE: 1, 10, 100, or 1000 events-per-click
RESPONSE:
FIXED: FAST response time adjustable from approximately 2 - 50
seconds. SLOW response is 5 times
fast setting
VARIABLE: Dependent on number of counts present. Typical times FAST
- 4 - 25 seconds,
SLOW - 4-60 seconds
(All times stated from 10% to 90% of final reading)
RATEMETER ALERT/ALARM: Set at any point corresponding to pre-selected
ratemeter range
SCALER ALARM: Adjustable from 1 - 999999 counts
SCALER COUNT TIME: Adjustable from 1 - 9999 seconds