Model 2401-EW Pocket Meter

PART #:48-2874


DETECTOR: 0.566" diameter end window G-M
SENSITIVITY: Typically 1050 cpm/mR/hr (137Cs gamma)
ENERGY RESPONSE: Energy dependent
METER DIAL: 0 - 2 mR/hr; 0 - 2.1k cpm; BAT OK
MULTIPLIERS: X1, X10, X100
LINEARITY: Reading within plus or minus 10% of true value
AUDIO: Built in unimorph speaker (select quiet position on main switch to silence the audio)
CALIBRATION CONTROLS: Accessible from front of instrument
RESPONSE: Typically 5 seconds from 10% to 90% of final reading
POWER: 1 ea. 9 volt
BATTERY LIFE: Typically 250 hours with alkaline batteries (battery condition can be checked on meter )
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint, and a recessed subsurface printed membrane front panel
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 1.8" (4.6 cm)H X 3.3" (8.4 cm)W X 5.3" (13.5 cm)L
WEIGHT: 0.9 lbs (0.4 kg) including batteries

Common Meter Dials

202-841
0-2 mR/hr; 0-2.1 kcpm

202-867
0-20 mSv/h; 0-2.1 kcpm

Accessories

Replacement Parts

Check Source
Other Accessories

Meter Movement
G-M Detector

Manual

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific    tel (310) 589 0601   fax (310) 919-2981,   Email

contact us: en espaņol, auf Deutsch
Open 8:00 AM - 5:00 PM Pacific Standard Time