The Model 133 series is a group of high range
energy compensated G-M detectors primarily used for area monitoring
applications
MODEL NUMBER
PART NUMBER
133-2
47-1717
133-2-1 WATERPROOF
47-2450
133-4
47-1674
133-4-1 WATERPROOF
47-2451
133-6
47-1718
133-6-1 WATERPROOF
47-2452
133-7
47-1216
133-7-1 WATERPROOF
47-2453
133-8
47-1226
133-8-1 WATERPROOF
47-2334
Note: The waterproof versions of the above detectors include a
hardwired 100' cable with "C" connector. Other lengths and connectors are
available on request.
INDICATED USE: Area monitoring DETECTOR: Halogen quenched G-M COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters,
and scalers, area monitors CONNECTOR: Series "C" (others available ) CONSTRUCTION: Anodized aluminum housing TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified to operate from -40° F(-40° C) to 150° F(65° C) SIZE: 1" (2.5 cm) diameter X 4" (10.2 cm)L WEIGHT: 0.2 lb (0.07kg)
133-2
133-4
133-6
133-7
133-8
LINEAR RANGE W/O DTC
0.1 mR/hr
200 mR/hr
1 mR/hr
2 R/hr
4 mR/hr
6 R/hr
25 mR/hr
30 R/hr
150 mR/hr
300 R/hr
LINEAR RANGE WITH DTC
0.1 mR/hr
1 R/hr
1 mR/hr
4 R/hr
4 mR/hr
10 R/hr
25 mR/hr
100 R/hr
150 mR/hr
1000 R/hr
SENSITIVITY
(cpm/mR/hr)
1000
(Cs-137 gamma)
100
(Cs-137 gamma)
30
(Cs-137 gamma)
4.2
(Co-60 gamma)
0.7
(Co-60 gamma)
ENERGY RESPONSE
(60 keV - 3 MeV)
Within plus or minus 25% of true value
Within plus or minus 15% of true value
Within plus or minus 25% of true value
DEAD TIME
Typically 80 microseconds
Typically 50 microseconds
OPERATING VOLTAGE
550 volts
460 volts
Response
Curves
Model
133-2 Response CurveModels
133-7 and 133-8 Response Curves