The Model 43-2-2 is a dual phosphor alpha/beta scintillator that is designed to be used for simultaneously counting alpha and beta contamination

PART NUMBER:47-2003


INDICATED USE: Alpha-beta survey
SCINTILLATOR: ZnS(Ag) adhered to 0.010" thick plastic scintillation material
WINDOW: Typically 0.8 mg/cm squared aluminized mylar (1.2 mg/cm squared recommended for outdoor use)
WINDOW AREA: Active and Open - 12 cm squared
EFFICIENCY (4pi geometry): Typically 30% - Pu-239; 30% - Th-230; 20% - Sr-90/Y-90; 5% - C-14
BACKGROUND:
Alpha - Less than 3 cpm
Beta - Typically 50 cpm or less (10 microR/hr field)
NON-UNIFORMITY: Less than 10%
CROSS TALK:
Alpha to Beta - Less than 10%
Beta to Alpha - Less than 1%
COMPATIBLE INSTRUMENTS: Model 2224, Model 2360, Model 2929
TUBE: 1.5"(3.8cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series "C" (others available)
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 2"(5.1cm) diameter X 7.3"(18.5cm)L
WEIGHT: 1 lb (0.5kg)

Replacement Parts

Mylar Window
Photomultiplier Tube
Scintillator

 

Instrumentation

Ludlum Dosimetry Shielding Portable Fume Hoods Specials

Hepa Vacuums

Air Sampling

Beakers

 Sources

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy SEARCH

Home

to order:    tel (888) 368 9118 or (310) 589 0601   fax (888) 440-1319,   Email

Direct Scientific / 124 San Tropez Ct. / Laguna Beach, California 92651  
contact us: en espaņol, auf Deutsch