The Model 44-142 is a beta
scintillator. With an active
area of 100 cm2 this detector
is good for surveys for beta contamination.
|
 |
| |
PART NUMBER: 47-3161 |
INDICATED USE:
Beta contamination survey
SCINTILLATOR: 0.01” thick plastic scintillator
WINDOW: 1.2 mg/cm2 recommended for outdoor use
WINDOW AREA:
Active - 100 cm2
Open - 89 cm2
EFFICIENCY (4pi geometry): Typically 4% - 14C; 30% -
90Sr/90Y; 20% - 99Tc
NON-UNIFORMITY: Less than 10%
BACKGROUND: Typically 300-350 cpm or less (10 mR/hr
field)
COMPATIBLE INSTRUMENTS: General purpose survey meters,
ratemeters, and scalers
TUBE: 1.125"(2.9cm) diameter magnetically shielded
photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series “C” (others available )
CONSTRUCTION: Aluminum housing with beige powdercoat
finish
TEMPERATURE RANGE: 5°F(-15°C) to 122°F(50°C)
May be certified to operate from -40°F(-40°C) to 150°F(65°C)
SIZE: 3.2"(8.1 cm)H X 3.5"(8.9 cm)W X 12.2"(31 cm)L
WEIGHT: 1 lb (0.5kg)
|
|
 |
|
Replacement Parts
Mylar Window
Photomultiplier Tube
Scintillator
Protective Cover
|
|