The Model 44-9-18 is a pancake G-M with a
telescoping handle and flexable neck that can be used with several
different instruments including survey meters, scalers, ratemeters, and
alarm ratemeters
PART NUMBER:47-2940
INDICATED USE: Alpha beta gamma survey; Frisking DETECTOR: Pancake type halogen quenched G-M WINDOW: 1.7 plus or minus 0.3 mg/cm² mica WINDOW AREA:
Active - 15 cm²
Open - 12 cm² EFFICIENCY(4pi geometry): Typically 5%-C-14; 22%-Sr-90/Y-90;
19%-Tc-99; 32%-P-32; 15%-Pu-239 SENSITIVITY: Typically 3300 cpm/mR/hr (Cs-137 gamma ) ENERGY RESPONSE: Energy dependent DEAD TIME: Typically 80 microseconds COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters,
and scalers OPERATING VOLTAGE: 900 volts CONNECTOR: Series "C" (others available ) CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 1.8" (4.6 cm)H X 2.7" (6.9 cm)W X 23" (58.4 cm)L; 36"(91.4cm)
fully extended WEIGHT: 1.5 lb (0.7kg)