Model 44-89 Square Cluster Pancake G-M

The Model 44-89 is a 4 pancake square cluster G-M detector for contamination survey.

PART NUMBER:47-2357

 

INDICATED USE: Alpha beta gamma survey
DETECTOR: 4 ea. pancake type halogen quenched G-M
WINDOW: 1.7 plus or minus 0.3 mg/cm² mica
WINDOW AREA:
Active - 63 cm²
Open - 50 cm²
EFFICIENCY(4pi geometry): Typically 5%-C-14; 23%-Sr-90/Y-90; 19%-Tc-99; 33%-P-32; 15%-Pu-239
SENSITIVITY: Typically 12,000 cpm/mR/hr (Cs-137 gamma )
ENERGY RESPONSE: Energy dependent
DEAD TIME: Typically 80 microseconds
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers
OPERATING VOLTAGE: 900 volts
CONNECTOR: Series "C" (others available )
CONSTRUCTION: Black ABS plastic with anodized aluminum back and stainless steel protective screen (79% open )
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 3.3" (8.4 cm)H X 5" (12.7 cm)W X 6.9" (17.5 cm)L
WEIGHT: 1.9 lbs (0.8kg)

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific    tel (310) 589 0601   fax (310) 919-2981,   Email

contact us: en español, auf Deutsch