 Model 44-9-18 Telescoping Pancake Detector

The Model 44-9-18 is a pancake G-M with a telescoping handle and flexable neck that can be used with several different instruments including survey meters, scalers, ratemeters, and alarm ratemeters

PART NUMBER:47-2940

 

INDICATED USE: Alpha beta gamma survey; Frisking
DETECTOR: Pancake type halogen quenched G-M
WINDOW: 1.7 plus or minus 0.3 mg/cm² mica
WINDOW AREA:
Active - 15 cm²
Open - 12 cm²
EFFICIENCY(4pi geometry): Typically 5%-C-14; 22%-Sr-90/Y-90; 19%-Tc-99; 32%-P-32; 15%-Pu-239
SENSITIVITY: Typically 3300 cpm/mR/hr (Cs-137 gamma )
ENERGY RESPONSE: Energy dependent
DEAD TIME: Typically 80 microseconds
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers
OPERATING VOLTAGE: 900 volts
CONNECTOR: Series "C" (others available )
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 1.8" (4.6 cm)H X 2.7" (6.9 cm)W X 23" (58.4 cm)L; 36"(91.4cm) fully extended
WEIGHT: 1.5 lb (0.7kg)