The Model 44-92 is a Xenon filled gas proportional detector for large area beta gamma survey. |
PART NUMBER:47-2362 |
| INDICATED USE: Low level beta-gamma survey WINDOW: 6.0 mg/cm² aluminized mylar (other thicknesses available) WINDOW AREA: Active - 169 cm² Open - 140 cm² EFFICIENCY (4pi geometry): Typically 10% - C-14; 13% - Tc-99; 24% - Cl-36; 26% - Sr-90/Y-90; 3% - I-129; 1.6% - Co-57; 1.2% - Fe-55; 5% - Pu-239 SENSITIVITY: Typically 36,000 cpm/mR/hr (Co-60 gamma) COMPATIBLE INSTRUMENTS: Model 12, Model 16, Model 18, Model 2000, Model 2200, Model 2221, Model 2241, Model 2241-2, Model 2241-3, Model 2350-1, Model 177-61 OPERATING VOLTAGE: Typically 1350 - 1600 volts COUNTER THRESHOLD SETTING: Typically 1.5 mV CONNECTOR: Series "C" (others available) CONSTRUCTION: Anodized aluminum housing TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 3.4"(8.6cm)H X 4.7"(11.9cm)W X 9.7"(24.6cm)L WEIGHT: 1.4 lbs (0.6kg) |
|
|
Direct Scientific tel (888) 368 9118 or (310) 589 0601 fax (310) 919-2981, Email |
|
|