Characteristics Under Test |
Range of Values of Influence Quantities |
Limits of Variation |
Pass / Fail |
|
AC Power |
102-132 VAC
178-238 VAC |
Reading cannot vary by more than plus or minus 5%
|
N/A |
|
Battery Power |
0 - 100 hours |
Reading cannot vary by more than plus or minus 10%
|
Pass |
|
Battery Power Indicator |
Test performed at the voltage that triggers the
battery failure indication |
Reading cannot vary by more than plus or minus 10%
|
Pass |
|
AC powered instrument with battery backup
|
Instrument must be marked for battery endpoint |
|
N/A |
|
Test performed at the voltage that triggers the
battery failure indication |
Readings cannot vary by more than plus or minus 10%
|
N/A |
|
Characteristics Under Test |
Range of Values of Influence Quantities |
Limits of Variation |
Pass / Fail |
|
Check Circuits |
Per manufacturer's recommendations |
|
|
|
Alarms (reset) |
Dose rate to activate alarm |
See section 5.2.1 |
N/A |
|
Alarms (delay) |
Dose rate to activate alarm |
Alarm must be indicated within 1 - 60 seconds |
N/A |
|
Alarm (threshold drift) |
Dose rate to activate alarm |
Alarm setpoint must not drift more than plus or
minus 10% over a 500 hour period |
N/A |
|
Stability |
3 hours (battery powered instruments) |
Reading cannot change by more than plus or minus 6%
|
Pass |
|
Stability |
24 hours (AC powered instruments) |
Reading cannot change by more than plus or minus 6%
|
N/A |
|
500 hours (AC powered instruments) |
Reading cannot change by more than plus or minus 15%
|
N/A |
|
Geotropism |
Tested in three mutually perpendicular orientations
|
Reading cannot vary by more than plus or minus 6%
|
Pass |
|
Response Time |
See Table 1 of Standard |
See Table 1 of Standard |
Pass |
|
Coefficient of Variation |
Greater than or equal to 1 mR/h, 1 mrd/h, 10 mrem/h,
2000 dpm |
Reading cannot change by more than plus or minus 10%
|
Pass |
|
Less than or equal to 1 mR/h, 1 mrd/h, 10 mrem/h,
2000 dpm |
Reading cannot change by more than plus or minus 15%
|
Pass* |
|
Line Noise Susceptibility |
See table 2 of standard |
Reading cannot change by more than plus or minus 15%
|
N/A |
|
Characteristics Under Test |
Range of Values of Influence Quantities |
Limits of Variation |
Pass / Fail |
|
Accuracy (photon dose rate) |
0.1 mrd/h - 1000 rd/h |
Cannot vary by more than plus or minus 15% of
conventionally true value |
NT |
|
Accuracy (count rate and contamination monitors)
|
50 dpm/square cm to 100,000 dpm/square cm |
Cannot vary by more than plus or minus 15% of
conventionally true value |
NT |
|
Accuracy (beta or neutron dose rate) |
0.1 mrem/h - 1000 rem/h |
Cannot vary by more than plus or minus 15% of
conventionally true value |
NT |
|
Probe surface sensitivity |
Stated by manufacturer |
|
NT |
|
Photon energy dependence |
80 keV - 1.25 MeV |
See equation in section 6.3 of standard |
NT |
|
20 keV - 3.0 MeV |
NT |
|
Beta Energy Dependence |
0.5 MeV - 3.5 MeV (Emax) |
See equation in section 6.3 of standard |
NT |
|
0.2 MeV - 3.5 MeV (Emax) |
NT |
|
Neutron Energy Dependence |
0.025 eV - 14 MeV |
See equation in section 6.3 of standard |
NT |
|
Photon Radiation Overload |
100X upper limit less than or equal to 10 rd/h |
Correct response within 2 minutes |
NT |
|
10X upper limit greater than 10 rd/h |
NT |
|
Angular Dependence |
0 - 45 degrees (photon) |
Instrument reading must not vary by more than plus
or minus 20% |
NT |
|
45 - 90 degrees |
Instrument reading must not vary by more than plus
or minus 50% |
NT |
|
0 - 45 degrees (beta) |
NT |
|
Characteristics Under Test |
Range of Values of Influence Quantities |
Limits of Variation |
Pass / Fail |
|
Temperature |
0 to 40 degrees C |
Reading cannot vary by more than plus or minus 15%
of reading at 22 degrees C |
Pass |
|
-10 to +50 degrees C |
Reading cannot vary by more than plus or minus 20%
of reading at 22 degrees C |
Pass |
|
10 to 35 degrees C |
Reading cannot vary by more than plus or minus 15%
of reading at 22 degrees C |
Pass |
|
Temperature Shock |
From -10% to 22 degrees C |
Pass |
|
From 50 to 22 degrees C |
Pass |
|
Humidity |
40 to 90% RH at 22 degrees C |
Readings cannot vary by more than plus or minus 15%
of the reading at 40% RH |
Pass |
|
Mechanical Shock |
50 g acceleration of 18 ms, half sine wave, test on
3 orthogonal axes (10 times) |
Reading cannot vary by more than plus or minus 15%
|
NT |
|
Vibration |
2 g acc., 10 - 33 Hz, test on 3 orthogonal axes for
15 min. |
NT |
|
Ambient Pressure |
70 - 106 kPa |
Pass |
|
Splashproof |
2 min. fine spray (4 L/min at 2 meters from nozzle)
|
Pass |
*Due to the relationship of the response time and the coefficient of
variation, readings on the lowest scale were taken using SLOW response
time (manufacturer's suggestion).