Fluke 10100AT TRIAD TnT Dosimeter Kit

10100AT TRIAD TnT Dosimeter Kit

The Model 10100AT TRIAD TnT Dosimeter Kit features the Model 35050AT Dosimeter, a technologically advanced, microprocessor-controlled, x-ray radiation dosimeter. The kit also includes ion chambers and test stand, triaxial/coaxial cable, AC adapter, HVL filter set, RS232 interface cable with adapters, customization software, instruction manual CD and lightweight carrying case.

 

Save time in setup and procedure

Expanded range of functionality


Product Specifications
Exposure and exposure rate accuracy:
  Basic accuracy of Model 35050AT is 1% of reading 2 range resolution steps over range of 18 to 28C and 2% of reading 2 range resolution steps over the full operating temperature range of 0 to 50C. 3% NIST-traceable calibration provided with each system and includes effects of Models 35050AT, 96035B and 96020C
Exposure time measurement:
Exposure time accuracy: 0.1% of reading 0.2 ms
Maximum exposure time: 6.5 s
Measurement resolution: 0.2 ms

Nominal measurement range:
100 to 20 Effective Range
Ion chamber: 15 cc
Units: R
Resolution step size: 1
100 to 20 Effective Range
Ion chamber: 15 cc
Units: R/sec
Resolution step size: 1
5 m to 1200 Effective Range
Ion chamber: 15 cc
Units: R/min
Resolution step size: 50
100 m to 72 k Effective Range
Ion chamber: 15 cc
Units: R/hr
Resolution step size: 1 m
2 to 333 m Effective Range
Ion chamber: 15 cc
Units: R/frame
Resolution step size: 0.02
1 to 175 m Effective Range
Ion chamber: 15 cc
Units: Gy
Resolution step size: 0.01
1 to 175 m Effective Range
Ion chamber: 15 cc
Units: Gy/sec
Resolution step size: 0.01
50 to 10.5 Effective Range
Ion chamber: 15 cc
Units: Gy/min
Resolution step size: 0.5
1m to 630 Effective Range
Ion chamber: 15 cc
Units: Gy/hr
Resolution step size: 0.01 m
0.02 to 2.9 m Effective Range
Ion chamber: 15 cc
Units: Gy/frame
Resolution step size: 0.5 2 n
10 to 2 Effective Range
Ion chamber: 150 cc
Units: R
Resolution step size: 0.1
10 to 2 Effective Range
Ion chamber: 150 cc
Units: R/sec
Resolution step size: 0.1
0.5 m to 120 Effective Range
Ion chamber: 150 cc
Units: R/min
Resolution step size: 5
10 m to 7.2 k Effective Range
Ion chamber: 150 cc
Units: R/hr
Resolution step size: 0.1 m
0.02 to 33 m Effective Range
Ion chamber: 150 cc
Units: R/frame
Resolution step size: 0.002
0.1 to 17.5 m Effective Range
Ion chamber: 150 cc
Units: Gy
Resolution step size: 0.001
0.1 to 17.5 m Effective Range
Ion chamber: 150 cc
Units: Gy/sec
Resolution step size: 0.001
5 to 1050 m Effective Range
Ion chamber: 150 cc
Units: Gy/min
Resolution step size: 0.05
0.1 m to 63.0 Effective Range
Ion chamber: 0.001 m
Units: Gy/hr
Resolution step size: 150 cc
0.001 to 290 Effective Range
Ion chamber: 150 cc
Units: Gy/frame
Resolution step size: 0.2 n
0.001 to 290 Effective Range
Ion chamber: 150 cc
Units: Gy/frame
Resolution step size: 0.2 n
2 to 2 Effective Range
Ion chamber: 150 cc VLDR
Units: R/sec
Resolution step size: 0.1
0.1m to 120 Effective Range
Ion chamber: 150 cc VLDR
Units: R/min
Resolution step size: 5
2 m to 7.2 k Effective Range
Ion chamber: 150 cc VLDR
Units: R/hr
Resolution step size: 0.1m
0.04 to 33 m Effective Range
Ion chamber: 150 cc VLDR
Units: R/frame
Resolution step size: 0.002
0.02 to 17.5 m Effective Range
Ion chamber: 150 cc VLDR
Units: Gy/sec
Resolution step size: 0.001
1 to 1050 m Effective Range
Ion chamber: 150 cc VLDR
Units: Gy/min
Resolution step size: 0.05
0.02 m to 63.0 Effective Range
Ion chamber: 150 cc VLDR
Units: Gy/hr
Resolution step size: 0.001 m
0.4 n to 290 Effective Range
Ion chamber: 150 cc VLDR
Units: Gy/frame
Resolution step size: 0.02 n
1 p to 100 n Effective Range
Ion chamber: Electrical units
Units: C
Resolution step size: 0.01 p
1 p to 100 n Effective Range
Ion chamber: Electrical units
Units: A
Resolution step size: 0.01 p
Note:
  Values for ion chambers are calculated using nominal sensitivities: 15 cc: 2.4 x 10 8 R/C 150 cc: 2.4 x 10 7 R/C Very Low Dose Rate effective range at 5% resolution steps. At 60 frames/sec (1 to 120 frames/selectable). IEC 61674 effective range at 1% resolution steps.
Measurement modes:
kVp/Dose/Time: Single shot, direct beam measurement of exposure, kVp, and time. Auto-ranging across three decade ranges. Auto reset between exposures. Display update after each exposure
kVp/Rate: Simultaneous measurement of kVp and exposure rate
Full sensitivity dose: Auto-ranging across five decades of sensitive ranges. Automatic drift & offset compensation, and automatic post-exposure display hold
Full sensitivity rate: Measurement range covers a span from low level image intensifier measurements to unattenuated direct beams. Automatic offset compensation and nonlinear filtering. Autoranging provides five decades of sensitivity ranges. Display updates once per second
Very Low Dose Rate (VLDR): This mode is only for making very low dose rate measurements. Nonlinear digital filtering and autoranging provide five decades of sensitivity ranges. Display updates once per second. In this mode, automatic current offset and drift compensation is disabled. As a result, the system can display very low dose rates
Power requirements:
  30-hour life with six AA alkaline batteries. Automatic power-down after user-selected period of unattended operation (5-255 minutes). An AC adapter is supplied with each 35050AT. When the AC adapter is in use, the auto power down feature is disabled, providing continuous operation. User selections for ion chamber, units, kV filter pack, temperature, pressure, and frame are stored in nonvolatile memory before automatic turnoff; eliminates manual re-selection at power-up
Bias voltage supply:
  Fixed electronic bias (~300 V). Bias voltage is removed from triaxial input connector at instrument turnoff
Customization:
  Allows user to modify contents of nonvolatile memory including ion chamber and kV filter pack conversion factors, temperature and pressure units, radiation units, and power down interval. A field customization software program is included for use with an IBM-PC or compatible
Display:
  Two-line, 20-character alphanumeric PLED (polymer LED), with 0.5 cm character height. Indicates all ion chamber/kV filter pack identification information, numerical measurement results, battery level, calibration date and other information
Weight:
  14 lb (6.4 kg)

Manual

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine