LAP GEMINI Room Laser

Picture of LAP GEMINI laser for patient alignmentThe GEMINI extended SSD treatment room laser is a unique LAP invention. We modified the standard ASTOR red laser and included an extra laser line 30 cm below the isocenter. This allows for extended SSD treatments and a lower patient setup.
 

 

 

-Dual isocenter
-Three laser lines
-Up to 30 cm extended isocenter range


Specifications:

Line width (up to 4 m distance) < 1 mm
 Line length (at 3 m / 10 ft distance) 3 m / 9 ft
 Laser type Diode
 Wavelength 635 nm
 Beam Color red
 Output power < 1 mW
 Laser class 2
 Power requirements 110/230 VAC, 5 VDC
 Main Voltage 5 VDC
 Power consumption 1.5 W
 Ambient temperature 0 - 40 °C
 Dimensions (H × W × D) 420 × 90 × 95 mm
 Weight 3 kg / 6.3 lbs
 Isocenter adjustment accuracy ± 0.5 mm
 Distance of horizontal lines 300 mm
subject to modifications
 
LAP patient alignment systems comply with Center of Devices and Radiological Health regulation for class II lasers (21 CFR 1040).