S.E. International Monitor 4 & 4EC Handheld l Radiation Alert® Detector - a ß Y X Monitor 4 pancake G-M detector by S.E. International

The MONITOR 4 and MONITOR 4EC are compact, general purpose survey meters capable of detecting alpha, beta, gamma, and x-rays over 3 selectable ranges. A red count light flashes and a beep sounds with each event detected. The Monitor 4EC, (energy compensated) offers a more accurate linear reading for gamma and x-rays (above 40 keV). Gamma sensitivity is typical of a pancake G-M detector: 1000 CPM/mR/hr (referenced to Cs-137). Energy Sensitivity for the 4EC is the Same as Monitor 4 except the energy response for gamma and x-rays through the detector sidewall is flat within +61% or -26% over the range of 40 keV to 100 keV, and within +35% or -17% over the range of 100 keV to 1.3 MeV.

 

Monitor 4 Pancake G-M Survey Meter    
Monitor 4EC Pancake G-M Survey Meter (Energy Compensated)    

Monitor 4 Survey Meter Specifications:

Detector
Monitor 4: Halogen-quenched uncompensated GM tube with thin mica window 1.5-2.0 mg/cm2 thick.
Monitor 4EC: Halogen-quenched GM tube, energy compensated sidewall 2mm tin filter. Thin mica window 1.5-2.0 mg/cm2 thick.
Display
Analog Meter Scale.
Monitor 4 Energy Sensitivity

Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%.
Detects beta at 50 keV with typical 35% detection efficiency.
Detects beta at 150 keV with 75% typical detection efficiency.
Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the sidewall of the detector.
Normal background is approximately 10-20 CPM.
Gamma Sensitivity:
1000 CPM/mR/hr (referenced to Cs-137)
Energy Sensitivity for the 4EC is the Same as Monitor 4 except the energy response for gamma and x-rays through the detector sidewall is flat within +61% or -26% over the range of 40 keV to 100 keV, and within +35% or -17% over the range of 100 keV to 1.3 MeV.

Operating Range

0-.5, 0-5, 0-50 mR/hr
0-500, 0-5,000, 0-50,000 CPM or 0-500 µSv/hr
0-50 mR/hr.

Range Switch
X1, X10, X100, Battery Check.
Accuracy
±15% of full scale (referenced to Cs-137)

S.E. International Monitor 4 Response Curve

Energy Response Curvey for the Monitor 4 Pancake G-M survey meter

S.E. International Monitor 4EC Response Curve

Energy Response Curvey for the Monitor 4EC Pancake G-M survey meter

Audio
Internally mounted beeper (can be switched off for silent operation).
Anti-Saturation
Meter will hold at full scale in fields as high as 100 times the maximum reading.

Power Requirements
One 9-volt alkaline battery. Battery life is up to 2,000 hours at normal background radiation levels
Temperature Range
-20° C to 55° C (-4° F to 131° F).
Weight
M4: 178 grams (6.3 oz.) without battery
4EC: 198.1 grams (7 oz.) without battery
Size
145 x 72 x 38 mm (5.7 x 2.8 x 1.5 in.)
Warranty
1 year limited warranty
Monitor 4 Kits
Build your own Monitor 4! Learn about how a geiger counter works!
Includes: All parts required to build a Monitor 4, Assembly and Illustration booklets.

Accessories

SI meter dial for Monitor 4 pancake G-M Survey meter

Carrying Case | Observer Software

Support

Monitor 4 & 4EC Operation Manual (pdf)| English | Español| France | Deutsch

Other SE International Survey Meters


Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific    tel (888) 368 9118 or (310) 589 0601   fax (310) 919-2981,   Email

contact us: en español, auf Deutsch