RAP Scintillation Probes - RAP47, RAP200, and RAP640 Scintillation Probes (gamma and x-ray)
The CsI (Tl) crystal used in Radiation AlertŪ Probes has a
higher atomic number, is more rugged, and less hygroscopic than a typical Nal
detector. The improved gamma ray absorption allows a thinner crystal to be used,
effectively reducing the background count rate.
Detector
Type-CsI (Tl) crystal (Cesium Iodide Thallium)
Probe Construction
Anodized aluminum housing. Approximate size: length 7-7.5 inches, diameter 1.5
inches. Connectors-Standard MHV. Cable (Optional)-Straight RG59/u, length 3.5
feet.
Temperature Range
-25°C to 40°C (-4°F to 104°F)
Humidity Range
0 to 90% Relative Humidity.
Weight
RAP47-Approx. 178 g (6.3 oz.) without cable
RAP200-Approx. 240 g (8.5 oz.) without cable
RAP640-Approx. 290 g (10.25 oz.) without cable
Warranty
One year limited.
| RAP47: Thin-crystal (1 mm) optimizes sensitivity to low energy gamma radiation while minimizing sensitivity to higher energies. An excellent detector for Iodine-125 and suitable for thyroid uptake measurements | RAP200: A 0.25-inch crystal optimizes this for the detection of special nuclear material (SNM) such as U-235, or related isotopes such as Am-241, much like a mini-FIDLER. |
![]() |
![]() |
| RAP640: Superior to a 1 x 1-inch NaI detector, this is a great choice for an all-around gamma and x-ray detector. Optimal energy range of about 20 keV to 2 MeV. Highest efficiency for a broad spectrum of gamma-emitting isotopes. |
All RAP Scintillation probes work with the
URSA-II
Universal Radiation Spectrum Analyzer. |
![]() |
|
|