Model 2363 Gamma/Neutron Survey Meter with Model 42-41 PRESCILA Probe

INDICATED USE: Gamma and Neutron Survey
SUGGESTED DETECTORS:
EXTERNAl: Model 42-41 PRESCILA detector
INTERNAL: Energy-compensated GM detector
SENSITIVITY:
Gamma: 1000 cpm per mrem/hr (internal detector)
;Neutron: 350 cpm per mrem/hr (with Model 42-41)
ADJUSTABLE ALARMS (Indicated by front panel - red LEDS):
Gamma alarm
Neutron alarm
Integrated dose alarm

LOGGING PUSHBUTTON: Located in the handle; used to store a reading
DATA LOGGER: Capable of logging up to1000 individual data points with the following identifiers for each point (All data is stored in non- volatile memory, allowing batteries to be removed without loss of data):
gamma and neutron sample counts
;sample number
date/time stamp
current integrated dose
10 character location identifier

RS-232 PORT: Located on the can, this allows the instrument to be connected to a PC for data download, and adjustment of setup parameters.
CONNECTOR: Series “C” (others available on request )
AUDIO: Built in unimorph speaker with volume control (greater than 60 dB at 2 feet, full volume )
AUDIO DIVIDE: Selectable dual or individual click-per-event for gamma and neutron counts
METER: 2.5"(6.4cm) arc, 1 mA analog type
METER DIAL: 0 - 10 mrem/hr, BAT OK, (overload) MULTIPLIERS: X0.1, X1, X10, X100
LINEARITY: Reading within 10% of true value DIGITAL DISPLAY: 6 digit LCD with 0.25" digits
SELECTOR SWITCH: Toggle switch to select gamma+neutron, gamma only, or neutron only
RESET/READ HV: A two position momentary action switch to allow for the meter to be reset or a reading of the current integrated dose on the digital display
RESPONSE: Will vary according to the number of counts present. Typically 2 - 11 seconds from
10% - 90% of final reading
POWER: 2 each “D” cell batteries (housed in sealed compartment that is accessible from front of instrument)
BATTERY LIFE: Greater than 150 hours (battery condition can be checked on meter )
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE:
GAMMA: -4° F(-20° C) to 122° F(50° C)
NEUTRON: 14° F(-10° C) to 122° F(50° C)
SIZE: 6.5"(16.5cm)H X 3.5"(8.9cm)W X 8.5"(21.6cm)L
WEIGHT: 3.5 lbs (1.6kg) including batteries

Prescila

Model 42-41 PRESCILA Probe

INDICATED USE: Neutron Survey
DETECTOR: PRESCILA Proton Recoil Scintillator
SENSITIVITY: Approximately 350 cpm per mrem/h
NEUTRON ENERGY RESPONSE: Thermal to 100MeV
ANGULAR DEPENDENCE: Within 15% over wide range of energies
GAMMA REJECTION: Approximately 500 cpm at 100mR/h (137Cs)
SUGGESTED INSTRUMENTS: Ludlum Models 12,  2363, 2241 series, and 2350-1
OPERATING VOLTAGE: Typically 500 - 700 volts
CONNECTOR: Series “C” (others available on request)
CONSTRUCTION: Aluminum housing w/black powdercoat finish
SIZE:10.125"(25.7cm)H X 4.25"(10.80 cm)W X 4.25"(10.80 cm)L
WEIGHT: 4.46 lbs.

 

 

Manual

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific  /  tel (310) 589 0601 /   Email

contact us: en español, auf Deutsch
Open 8:00 AM - 5:00 PM Pacific Standard Time
Follow us on twitter and youtube.
visit our youtube channel
follow us on twitter
Check out our instrument demos.