The Model 44-21 is a dual crystal scintillation detector utilizing a thin plastic scintillator and 1" X 1mm NaI(Tl) crystal for beta and gamma measurements.

PART NUMBER:47-1560

INDICATED USE: Beta, low energy gamma detection
SCINTILLATOR: 1" (2.5 cm) diameter X 1 mm thick NaI(Tl) crystal coupled to a 1" (2.5 cm) diameter X 0.01" (0.03 cm) thick plastic scintillator
WINDOW: 1.2 mg/cmē aluminized mylar
WINDOW AREA: Active and Open - 5.1 cmē
EFFICIENCY(4pi geometry): Typically 19% - I-125; 5%-C-14; 28%-P-32
SENSITIVITY: Typically 40 cpm/microR/hr (Cs-137 gamma )
ENERGY RESPONSE: Energy dependent
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers
TUBE: 1.5:(3.8cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series "C" (others available )
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 2" (5.1 cm) diameter X 7.3" (18.5 cm)L
WEIGHT: 2 lb (0.9kg)

 

Replacement Parts

Photomultiplier Tube
Scintillator
Protective Cap

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific  /  tel (310) 589 0601 /   Email

contact us: en espaņol, auf Deutsch
Open 8:00 AM - 5:00 PM Pacific Standard Time
Follow us on twitter and youtube.
visit our youtube channel
follow us on twitter
Check out our instrument demos.