Model 177 Alarm Ratemeter

  • 4 Ranges
  • Adjustable Alarm
  • Utilizes G-M and Scintillation Detectors
  • Battery Back-up
  • Pulse Output
  • Recorder Output
  • Total Counting Range from 0 - 500,000 cpm

PART NUMBER:48-1632

INDICATED USE: Multi-purpose radiation monitoring; frisking
COMPATIBLE DETECTORS: G-M, scintillation
METER DIAL: 0 - 500 cpm, 0 - 1.5 kV, BAT TEST (others available )
MULTIPLIERS: X1, X10, X100, X1k
LINEARITY: Reading within plus or minus 10% of true value with detector connected
CONNECTOR: Series "C" (others available )
AUDIO: Built in unimorph speaker with volume control (greater than 60 dB at 2 feet full volume )
ALARM: Indicated by a red lamp and full volume audible tone
DATA OUT: 9 pin series "D" connector allowing for external connections of the following items: battery; unregulated supply; instrument common ground; alarm signal (current sink); pulse out; external background subtract; and recorder output.
RECORDER OUTPUT: Correlated to meter reading (adjustable to 1 volt at 1 mA)
CALIBRATION CONTROLS: Accessible from back of instrument (protective cover provided )
HIGH VOLTAGE: Adjustable from 200 - 1500 volts (can be read on meter )
THRESHOLD: Adjustable from 10 - 100 mV
RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
RESET: Push-button to zero meter
POWER: 95 - 135 VAC (178 - 240 VAC available), 50 - 60 Hz single phase (less than 100 mA), 6 volt sealed lead acid rechargeable battery (built in)
BATTERY LIFE: Typically 50 hours (battery condition can be checked on meter )
BATTERY CHARGER: Battery is continuously trickle charged when instrument is connected to line power and turned on (optional external fast charger is available)
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 5" (12.7 cm)H X 8" (20.3 cm)W X 6" (15.2 cm)L
WEIGHT: 4.2 lbs (1.9 kg)

 
*For Mylar window scintillators it is strongly suggested that the optional overload indicator be used to better determine the presence of a light leak.

Common Meter Dials

 202-096
0 - 500 cpm; 0 - 1.5 kV
For Any Detector

 202-642
0-660 cpm; 0-0.2 mR/hr
For Model 44-9; 44-40; 44-88; 44-89; 44-94

 202-704
0-660 cpm; 0-2
mSv/h
For Model 44-9; 44-40; 44-88; 44-89; 44-94

 


 202-420
0-420 cpm; 0-0.2 mR/hr
For Model 44-7
 

202-861
0-420 cpm; 0-2 mSv/h
For Model 44-7

202-698
0-0.5
mR/hr
For Model 44-2

 

Accessories

Replacement Parts

Check Source
Overload Indicator
Relay Output
Recorders
Battery Charger

Cable
Meter Assembly
Meter Movement
Battery

 

 

Nuclear Instrumentation

Dosimetry Shielding Sources Specials

Hepa Vacuums

Chemical Detection

Marinelli Beakers

Fume Hoods

Accessories

Emergency Response

Nuclear Medicine

Radiation Therapy Search Home

Direct Scientific  /  tel (310) 589 0601 /   Email

contact us: en espaņol, auf Deutsch
Open 8:00 AM - 5:00 PM Pacific Standard Time
Follow us on twitter and youtube.
visit our youtube channel
follow us on twitter
Check out our instrument demos.