Dose-I Electronic Personal Dosimeter

  • Model: Dose-i

$495.00

Add to Cart:
Sieverts ONLY in stock.

Displays dose and dose rate
- Water and noise resistant
- Large, clear display and alarm functions
- Pocket-sized and lightweight
- Standard Disk battery for ease of replacement

The Electronic Personal Dosimeter DOSE-I is an effective response to gamma radiation. The energy range of the DOSE-I is spans 35 keV to 3 MeV. The DOSE-I features measurement of personal dose and dose rate of gamma - X-Ray radiation exposure, providing visual and audible Dose / Dose Rate alarms. The DOSE-I can be configured in Sv and rem units through the Dosimeter Configuration Software and IR reader.

Specifications

1. Type: DOSEi-γ for gamma (X) rays

2. Detector: Silicon semiconductor detector

3. Radiation Detected: Detected Gamma (X) rays

4. Display Range:

0.001 mSv to 999.9 mSv
0.001 mSv/h to 999.9 mSv

5.Accuracy: ≤ ±10% (0.01 to 999.9 mSv, 137Cs ref.) / ≤ ±10 % (1.0 mrem to 99.99 rem, 137Cs ref.)

6. Linearity of doseequivalent rate: ≤ ±10 % (up to 100 mSv/h) / ≤ ±10 % (up to 10 rem/h)

7. Energy Range: 35 keV to 3 MeV

8. Display: 4 digit liquid display (0.001 to 9999), switching and indicating function of dose rate low battery indication, overflow indication when exceeding measurement range

9. Alarm Setting

Settable for every 1 mrem, 1 mrem/h (setting with software)
Settable for every 0.01 mSv, 0.01 mSv/h (setting with software)

10. Alarm Output: Alarm sound (large, medium, small, OFF sound volume settable), alarm lamp blinking

11. Data Storage: Up to 600 trend data records (for only gamma rays)*, with non-volatile memory

12. Test Function: When switch is ON, operation check is available with alarm output and liquid display

13. Operating Humidity: ≤ 90 % RH (non condensing)

14. Power Source: Coin-type lithium battery (CR2450)

15. Battery Life: 720 hours or more (without alarm indication)

16. Size

Approx. 30 (W) x 110 (H) x 12 (D) mm
Approx. 1.8 (W) x 4.3 (H) x 0.47 (D) in.

17. Mass: Approx. 60 g / Approx. 0.13 lb

18. Communication: Infrared communication (communication cable and software are optional)

19. Other: Water resistant (JIS protection grade 4), noise resistant (mobile phones ,etc.)

*Utilizing trend function requires infrared communication cable and software (optional).

CAUTION: Read the instruction manual provided before using this product, to make sure you operate it safely.

This product was added to our catalog on Sunday 22 September, 2019.