MODEL 2241-2RK Nuclear Emergency Response Kit
This kit includes a general purpose survey
& gross counting meter with G-M and scintillation detectors. A locking
toggle switch allows for quick change between two detector setups. Alert/Alarm
is indicated by annuciator on display and audible tone. Auto ranging from
0.0 mR/hr - 9999 R/ hr; 0.000 mSv/hr - 9999 Sv/hr; 0 cpm - 999k cpm; or
0 cps - 100k cps. Scaler is activated by push button in handle (count time
adustable from 1 - 9999 seconds in 1 second time intervals). Battery life
is typically better than 200 hours and scaler alarm is adjustable from 1
- 999999 counts. Dose rate alarm is adjustable at any point corresponding
to pre-selected ratemeter range.
Model 44-9 (Included in Kit)
Includes Alpha, Beta & Gamma pancake detector. The efficiency (4pi
geometery) is typically 5%-14C; 22%-90Sr/90Y; 19%-99Tc; 32%-32P; 15%-239Pu.
The sensitivity is typically 3300 cpm/mR/hr (137Cs).
Model 44-2 (Included in Kit)
High energy gamma detector for MICRO-R sensitivity. Scintillator is 1
diameter X 1 thick NaI(Tl) crystal. Sensitivity is typically 175,000
cpm/mR/hr (Cs-137).
Model 10362 Case
(Included in kit) Great for emergency vehicles!
The pelicase is an air/water tight foam fitted case. Inside dimensions are
17L x 11.7W x 6D.
Cs-137 Check Source and Holder (Included in kit) This exempt quantity 1
uCi check source and holder is mounted on the side of the instrument and
used to check for proper instrument operation in the field.
If special calibration requirements are not initially
specified, calibration will be made based on the default settings shown
below:
| |
Color Dot:
Probe model:
Units/Timebase:
Rate Alarm:
Rate Alert:
Scaler Alarm:
Scaler count time: |
|
Red
44-9
cpm
50 kcpm
20 kcpm
85000
12 sec. |
Yellow
44-2
R/hr
50 µR/hr
20 µR/hr
85000
12 sec. |
|
|
|
|
|
|
|
The following parameters are adjusted by removable switchboard. Once
initial setup and calibration have been completed the switchboard can be
removed without affecting normal operation of the instrument (Model
2241-2 may be ordered without switchboard)
BACKLIGHT "ON" TIME: 5, 30, 60, 90, 180, or 240 seconds for the
backlight to stay "ON" when activated.
SET MINIMUM DISPLAY: Allows lower limit of auto-ranging display to
be fixed. For example, display can be set to not show values less than 1
R/hr.
RS-232 DATA DUMP MODE: Enables or disables dump mode to RS-232
port. When enabled data will be dumped every 2 seconds
RS-232 DETECTOR SETUP MODE: Allows for detector parameters to be
input by RS- 232 port
BAUD RATE: 150, 300, 600, 1200, 2400, 4800, 9600, or 19200 bps.
DEAD TIME: Adjustable from 0 - 9999 seconds
CALIBRATION CONSTANT: Adjustable from 0.001 - 280 X 109
counts/display unit
DISPLAY UNITS: Can read in R/hr, Sv/hr, cpm, or cps
TIME BASE: Can read in seconds or minutes
AUDIO DIVIDE: 1, 10, 100, or 1000 events-per-click
RESPONSE:
FIXED: FAST response time adjustable from approximately 2 - 50
seconds. SLOW response is 5 times fast setting
VARIABLE: Dependent on number of counts present. Typical times FAST
- 4 - 25 seconds, SLOW - 4-60 seconds
(All times stated from 10% to 90% of final reading)
RATEMETER ALERT/ALARM: Set at any point corresponding to
pre-selected ratemeter range
SCALER ALARM: Adjustable from 1 - 999999 counts
SCALER COUNT TIME: Adjustable from 1 - 9999 seconds |
|
Model 44-9 Pancake Probe
|
|
|
INDICATED USE: Alpha beta gamma survey; Frisking
DETECTOR: Pancake type halogen quenched G-M
WINDOW: 1.7 plus or minus 0.3 mg/cm² mica
WINDOW AREA:
Active - 15 cm²
Open - 12 cm²
EFFICIENCY(4pi geometry): Typically 5%-C-14; 22%-Sr-90/Y-90;
19%-Tc-99; 32%-P-32; 15%-Pu-239
SENSITIVITY: Typically 3300 cpm/mR/hr (Cs-137 gamma )
ENERGY RESPONSE: Energy dependent
DEAD TIME: Typically 80 microseconds
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters,
and scalers
OPERATING VOLTAGE: 900 volts
CONNECTOR: Series "C" (others available )
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 1.8" (4.6 cm)H X 2.7" (6.9 cm)W X 10.7" (27.2 cm)L
WEIGHT: 1 lb (0.5kg) |
|
Model 44-2 Pancake Probe
|
|
|
INDICATED USE: High energy gamma detection
SCINTILLATOR: 1" (2.5 cm) diameter X 1" (2.5 cm) thick sodium
iodide (NaI)Tl scintillator
SENSITIVITY: Typically 175 cpm/microR/hr (Cs-137)
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters,
and scalers
TUBE: 1.5:(3.8cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series "C" (others available )
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 2" (5.1 cm) diameter X 7.3" (18.5 cm)L
WEIGHT: 1 lb (0.5kg) |
|
|
Model 2241-2 Display
|
|
2241-2RK |
Model 2241-2RK with Model 44-9, Model 44-2 |
|
|